Specifications - Barnant Company 877-9605 Operating Manual

Liquid-sensor controllers, liquid-pad detectors
Table of Contents

Advertisement

Output:
Contact Closure Output Rating: 30V DC—1 Ampere Maximum
Audible Alarm Sound Level:
Input:
Line Voltage Limits
(115V AC Controller):
(230V AC Controller):
Power Dissipation:
AC Line Current:
Installation Category:
Construction:
Controller Dimensions:
LIQUID-PAD
Detector Dimensions:
Weight:
Detector Cable Length:
Enclosure Rating:
Environment:
Operating Temperature Range: 0°C (32°F) to 40°C (104°F)
Storage Temperature:
Humidity (non-condensing):
Barometric Pressure:
Altitude:
Pollution Degree:
Chemical Resistance:
LIQUID-SENSOR Controller:
LIQUID-PAD Detector:
Compliance:
* Unit operating voltage range as tested by Underwriters Laboratories. This unit has been designed and tested by the
manufacturer for an extended operating voltage range of 90–132V AC.

SPECIFICATIONS

90 dBA
102–132V AC 50/60 Hz Single Phase*
190–260V AC 50/60 Hz Single Phase
1.2 kVA maximum
10 A max at 115V AC
6.3 A max at 230V AC
Installation Category II per IEC 664
(Local level—appliances, portable
equipment, etc.)
3 in H
5.0 in
3 lbs
3.5 feet (others available)
IP22 per IEC 529
−45°C (−49°F) to 65°C (149°F)
10 to 85%
75 to 106 kPa
less than 2000 m
Pollution Degree 2 per IEC 664
(Indoor Usage—lab, office)
All materials withstand common
cleaning solvents. Construction
materials used are polyester
labels, ABS plastic enclosure and
zinc-plated screws.
Top layer — Flat polyester attached
with polyester laminate.
Bottom layer — Polyester attached
with polyester laminate.
Connector is ABS plastic.
115V: UL508, CSA C22.2, No. 14-M91
230V (For CE Mark):
EN61010-1/A2: 1995
(EU Low Voltage Directive) and
EN61326-1/A1: 1998 (EU EMC Directive)
20
8.1 in W
6.3 in D
2.6 in

Advertisement

Table of Contents
loading

This manual is also suitable for:

877-9600877-9500877-9501

Table of Contents