Maintenance And Troubleshooting; Fuse Replacement; Cleaning - Barnant Company 877-9605 Operating Manual

Liquid-sensor controllers, liquid-pad detectors
Table of Contents

Advertisement

the sensitivity control should be set for proper operation. The last column
indicates the quantity of fluid required in the target area of the detector to
activate the detector.
Sensitivity will change with detector position or when used near metallic
objects (e.g., table). Also, sensitivity changes when using longer (optional)
cables.
CAUTION:
Use with caus tic liq uids not com pat i ble with de tec tor
materials voids war ranty.
Fluid Type
Distilled Water
Tap Water
Acetic Acid (10% max.)
Alcohols
Glycerin
10W Motor Oil

MAINTENANCE AND TROUBLESHOOTING

WARNING: No user ser vice able parts are in side of this Con trol ler.
Refer ser vic ing to your dealer.

Fuse Replacement

The 115V AC LIQUID-SENSOR Controller uses a 10-A Slo-Blo fuse (3AG,
part number B-1115-0052), which is located in the fuse holder on the rear
panel. The 230V AC LIQUID-SENSOR Controller uses a 6.3-A fuse (5 x 20
mm, part number B-1115-0054), which is located in the drawer in the
POWER INPUT connector on the rear panel. Replace only with the correct
fuse.

Cleaning

Clean the top surface of the detector pad with a soft, dry cloth. If liquid cleaner
is used, rinse and dry thoroughly before resuming operation. Keep the
LIQUID-SENSOR Controller enclosure clean by using a mild detergent.
Never immerse or use excess fluid.
Table 1. Typical Sensitivity Settings
Setting (Turns ccw
from activation point)
1
1
2
1
1
2
1
1
2
1
1
2
1
1
2
1
2
17
Nominal Volume
< 1 mL
< 1 mL
< 0.5 mL
< 0.5 mL
< 2.0 mL
< 2.0 mL