NOTE: Retain all packing materials until proper operation has been verified.
WARNING: The LIQ UID-SENSOR Con trol ler must be po si tioned so
leak ing liq uid will not en ter the rear power con nec tors. Use
of a power line with a Ground Fault In ter rupt (GFI) is rec om -
mended.
Proceed as follows to setup the LIQUID-SENSOR Controller:
1. Place the rear panel power OFF-ON switch in the OFF position.
NOTE: For 115V AC pump models, the pump power cord will plug directly
into the MAIN OUTPUT or AUXILIARY OUTPUT connectors on the
rear panel of the LIQUID-SENSOR Controller. For 230V AC models,
use pump cable 77096-50 for pumps with European connectors or
pump cable 77096-55 for pumps with other style connectors.
2. Connect pumps as follows:
a. If a main pump and a standby (auxiliary) pump are to be used,
connect the main pump to the MAIN OUTPUT connector on the rear
panel and connect the standby (auxiliary pump) to the AUXILIARY
OUTPUT connector.
b. If only the main pump is to be used with no standby, and the pump is
to shut down when an alarm condition exists, connect the main
pump to the MAIN OUTPUT connector. The AUXILIARY OUTPUT
connector can be used to power other devices such as lights, remote
alarms or other equipment.
c. If only the main pump is to be used with no standby, and the pump is
not to shut down when an alarm condition exists, connect the main
pump to a separate power source. The MAIN OUTPUT connector
and the AUXILIARY OUTPUT connector can be used to power other
devices such as lights, remote alarms or other equipment.
3. Connect one end of the detector cable(s) to the front panel INPUT
connector(s) and the other end to the desired detector(s). Only one
detector can be connected to each INPUT connector. Detectors must be
connected using a cable configured as shown in APPENDIX A. Make
sure cable connections are secure or unit will not function
properly.
4. The LIQUID-PAD Detector must be positioned under the pump so that
any leakage will drip onto the detector pad target area (area outlined
with the black lines). The detector can be secured in place using
hook-and-loop fabric fasteners or can be permanently fastened with #4
screws. Two circular marks are located on the outer edge of the detector

INSTALLATION AND SETUP

9