12.1.2. Data Format

There are two type of data format selected at F-Function f-45.
The standard serial data output (SER.OUT) uses format 1.
Format 1
Format 2
Header 1
Stable weighing data.
ST
Unstable weighing data.
US
Overload (Out of range).
OL
Header 2
GS
G~
or
NT
N~
or
TR
T~
or
PT
Data
The first of the data bits is the polarity whether "+" or "-".
When data is zero, the polarity is "+".
Format 1 is 8 figures (bits) including polarity and decimal point(dot or comma).
Format 2 is 9 figures (bits) including polarity and decimal point(dot or comma).
In case of "out of range", the data bits of format 1 are replaced by spaces except the
decimal point.
In case of "out of range", the data bits of the format 2 are replaced by all 9's
Unit
Function Cf-10 is used in common with RS-232C and standard serial data output.
In case of Cf-10 0 or 2, the unit length is 2 figures.
In case of Cf-10 1 or 3, the unit length is 3 figures. Depending on circumstances,
an A&D printer may not work correctly.
AD-4329A Weighing Indicator
Gross data.
Net data
Tare data.
Preset tare data.
Page 45