6.2. F-Functions

Digital Filter
Item
f-00
Filter
Motion / Averaging
time
f-01
Zero tracking
f-02
MD mark
Motion detection
condition
d: division (weighing interval) of first range.
s: second.
*: factory settings (Initial settings).
AD-4329A Weighing Indicator
Parameter
0
2 d/ 1.6s
1
4 d/ 1.6s
2
8 d/ 1.6s
3
16 d/ 1.6s
4
32 d/ 1.6s
5
64 d/ 1.6s
6
128 d/ 1.6s
7
2 d/ 3.2s
* 8
4 d/ 3.2s
9
8 d/ 3.2s
10
16 d/ 3.2s
11
32 d/ 3.2s
12
64 d/ 3.2s
13
128 d/ 3.2s
0
OFF
1
0.5 d/ 1s
2
1.0 d/ 1s
3
1.5 d/ 1s
4
2.0 d/ 1s
5
2.5 d/ 1s
6
0.5 d/ 2s
7
1.0 d/ 2s
* 8
1.5 d/ 2s
9
2.0 d/ 2s
10
2.5 d/ 2s
0
No motion detection
1
0.5 d/ 0.5s
2
1.0 d/ 0.5s
3
2.0 d/ 0.5s
4
3.0 d/ 0.5s
5
4.0 d/ 0.5s
6
0.5 d/ 1s
7
1.0 d/ 1s
* 8
2.0 d/ 1s
9
3.0 d/ 1s
10
4.0 d/ 1s
Page 27
Description
If weak filter is set, the
response will be fast, but will
be more sensitive to external
influences such as vibration.
This function traces the
weighing value drifting around
zero point slowly, displayed as
zero. If a strong parameter is
set, a very small zero drift may
be not detected.
If Cf-03 is 1, the 0, 1, 6 or 7
can be selected only.
If Cf-03 is 1, the initial setting
sets 7.
The function to set the
condition of judgment whether
a weighing value is unstable or
stable. The
ZERO
key has effective in the
TARE
stable state. If these keys need
in the unstable state, set to
f-02 0. In case of Cf-03 1,
f-02 6 or f-02 7 can be
selected only. (Internal setting
is 6.)
key and
6. Functions

Hide quick links:

Advertisement