2. Common operations
1)
On each measurement screen, you can toggle between "Data List" display and "Trend Graph" display.
2)
Click
to start the measurement with the settings applied.
3)
Click [Save to CSV] to save the data to a file. Click [Clear] to clear data.
* The data is saved in CSV format.
4)
"Right-click" functions
Right-click on the data list or graph to display a pop-up menu from which you can copy the data or graph to the
clipboard.
<Example>
You can also specify the background color (B/W) and size for the graph that you want to save.
3. Individual functions
Zero
calibration
Zero calibration [0-CAL]
• When you click [0-CAL], the shutter inside the probe automatically closes, and zero calibration starts.
• It is recommended that zero calibration be used after start-up or whenever there is a change in ambient temperature or a
temperature change inside the instrument due to using it for long periods.
Settings [Settings]
• The pop-up window for setting the measurement conditions appears.
• You can set the measurement speed and the calibrated channel.
Data import [Read Data]
• This function imports the data from the data processor.
User calibration [User Cal] (only for chromaticity measurement)
• The pop-up window for user calibration appears. (See User calibration on P.60)
• The calibration channel (CH00 to CH99) data is read before displaying the pop-up window.
Gamma measurement [Gamma] (only for chromaticity measurement)
• The pop-up window for gamma measurement appears. (See 3-1. Gamma measurement (only for chromaticity
measurement) on P.50.)
Settings
Data import
49
User
Gamma
calibration
measurement