Download Print this page

ResMed Compact Bilevel Device VPAP Auto Brochure & Specs page 2

Resmed compact bilevel device brochure

Advertisement

VPAP AUTO FEATURES SUMMARY
Modes of operation
• VAuto
• S (Spontaneous)
• CPAP
Product features
• Simplified standard setting menu
• Vsync
automatic leak management
• TiControl
(Ti max/min)
• Adjustable breath trigger/cycle
• Adjustable exhalation comfort
• 12V DC input (via converter)
• Auto-adjusting bilevel pressures
• Compact S8 design
• AutoSet technology
• Smooth pressure delivery
Heated humidification
• Optional, fully integrated
Data storage
• Usage and Treatment Summary (day, week, month, six-month, year)
• Usage profile
• Summary data storage
• Detailed data storage
• Remote settings adjustment
Data transfer
• Direct download to PC
• ResLink via SmartMedia card
• ResScan Data Card
ResMed Corp Poway, CA, USA +1 858 746 2400 or 1 800 424 0737 (toll free), ResMed Ltd Bella Vista, NSW, Australia +61 (2) 8884 1000 or 1 800 658 189 (toll free). Offices in Austria, Brazil, Finland, France, Germany,
Hong Kong, Japan, Malaysia, Mexico, Netherlands, New Zealand, Singapore, Spain, Sweden, Switzerland, United Kingdom (see website for details). ResControl, ResLink, ResScan, ResTraxx, S8, TiControl, VPAP Adapt
SV and Vsync are trademarks of ResMed Ltd. AutoSet, HumidAire and VPAP are trademarks of ResMed Ltd and registered in U.S. Patent and Trademark Office.
Global leaders in sleep and respiratory medicine
TECHNICAL SPECIFICATIONS
Performance
Operating pressure range: 3 to 20 cm H
Maximum single fault pressure: 30 cm H
Dynamic pressure characteristics
VAuto mode: 4 to 20 cm H
O (Pressure support: 0 to 10 cm H
2
S mode: IPAP: 5 to 20 cm H
O; EPAP: 3 to 20 cm H
2
CPAP mode: 4 to 20 cm H
O
2
Sound pressure level
< 24 dB (tested in accordance with the requirements of ISO 17510-1:2002)
Data storage
Data includes usage, leak, pressure, AHI, AI, tidal volume, respiratory rate
and minute ventilation
Via LCD menu: 365 sessions of summary data
Via ResLink: 365 sessions of summary data, 30 sessions of detailed data
Via ResScan Data Card: 180 sessions of summary data, two sessions of
detailed data
Dimensions (L x W x H): 4.4" x 6.5" x 5.7" (112 mm x 164 mm x 145 mm)
Weight: 2.9 lb (1.3 kg)
Power supply
Input range 100–240V, 50–60Hz, 40VA (typical power consumption),
< 100VA (maximum power consumption)
Environmental conditions
Operating temperature: 41°F to 97°F (+5°C to +36°C)
Operating humidity: 10–95% non-condensing
Electromagnetic Compatibility
Product complies with all applicable electromagnetic compatibility
requirements (EMC) for residential, commercial and light industry
environments according to IEC 60601-1-2.
Air filter
Two-layered, powder-bonded, polyester non-woven fiber
IEC 60601-1 classification
Class II (double insulated), Type CF
Note: The manufacturer reserves the right to change these specifications
without notice.
PRODUCT CODES
VPAP Auto
HumidAire 3i
©2007 ResMed. 1012117/2 07 12
www.resmed.com
O
2
O
2
O)
2
O
2
26301
33906

Advertisement

loading

This manual is also suitable for:

Vpap auto