Technical Specifications

Operating pressure range
Maximum single fault steady
state pressure
Pressure measurement
tolerance
Flow measurement tolerance
VAuto mode
S, ST and T modes
CPAP mode
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871:
Sound pressure level
Sound power level
Dimensions (L x W x H)
Weight
Power supply
Housing construction
Operating temperature
Operating humidity
Storage and transport
temperature
Storage and transport humidity
Operating altitude
Supplemental oxygen
Electromagnetic compatibility
Air filter
Air tubing
Air outlet
IEC 60601-1 classification
Note: The manufacturer reserves the right to change these specifications without notice.

Symbols Which Appear On The Device

Attention, consult accompanying documents;
Dangerous voltage;
3 to 25 cm H
O
2
40 cm H
O
2
±0.5 cm H
O ± 4% of the measured reading
2
±0.1 or 20% of reading, whichever is greater
4 to 25 cm H
O (measured at the mask); Min EPAP: 4 cm H
2
Max IPAP: 25 cm H
O; Pressure Support: 0 to 10 cm H
2
IPAP: 4 to 25 cm H
O (measured at the mask); EPAP: 3 to 25 cm H
2
(measured at the mask); Pressure Support: 0 to 22 cm H
4 to 20 cm H
O (measured at the mask)
2
26 dBA with uncertainty of 2 dBA as measured according to
ISO 17510-1: 2002
28 dBA with uncertainty of 2 dBA as measured according to
ISO 17510-1:2007
36 dBA with uncertainty of 2 dBA as measured according to
ISO 17510-1:2007
4.4" x 6.5" x 5.7" (112 mm x 164 mm x 145 mm)
2.9 lb (1.3 kg)
Input range 100–240V, 50–60Hz, 40VA (typical power consumption),
< 100VA (maximum power consumption)
Flame retardant engineering thermoplastic
41 º F to 95 º F (+5 º C to +35 º C)
10–95% non-condensing
-4 º F to 140 º F (-20 º C to +60 º C)
10–95% non-condensing
Sea level to 8,500' (2,600 m)
Recommended maximum supplemental oxygen flow:
4 L/min (VAuto mode); 15 L/min (CPAP , S, ST and T modes)
Product complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC60601-1-2, for residential,
commercial, and light industry environments
Two-layered, powder-bonded, polyester non-woven fiber
Flexible plastic, 1 x 6'6" (2 m)
The 22 mm conical air outlet complies with ISO 5356-1
Class II (double insulation), Type CF
Class II equipment;
Drip proof;
Type CF equipment;
Start/Stop;
Manufacturer.
Technical Specifications
O;
2
O
2
O
2
O
2
15