Fluke 975CK Instruction Sheet page 2

Gas sensor calibration kit

Advertisement

Calibrating the CO and CO
Calibrate the CO and CO
To simultaneously calibrate both sensors,
If not already in calibration mode, press L, R, and
1.
S simultaneously for 3 seconds to enter calibration mode.
Press 1
2.
Press 3
3.
The Meter display reads:
Calibration Procedure
Apply Nitrogen...
4.
Apply ½ liter/minute of nitrogen for 2 minutes. Press
[Cancel]
3
5.
When sensor zero is complete, remove the nitrogen and
press 2
The Meter displays:
Apply CO Calgas concentration...
Enter the concentration amount using 1[DOWN] and
6.
2[UP] to change the entry and 3[ENTER]to store it.
7.
Apply the calibration gas mixture of CO and CO
2[OK] or 3[Cancel] to exit calibration.
8.
When finished, remove gas and press
to next calibration date. Use 1[DOWN] and 2[UP] to
change the number and 3[ENTER]to store the number.
The calibration procedure is complete.
Calibrating the CO Sensor
To calibrate the CO sensor:
With the calibration cap on, press L, R, and S
1.
simultaneously for 3 seconds to enter calibration mode.
Press 1 [Yes] to initiate the calibration procedure or
2.
press 3[Cancel] to exit calibration mode.
Press 1[CO]. The Meter zeros the CO sensor. If desired,
3.
press 3[Cancel] to abort Meter calibration.
Use 1[DOWN] and 2[UP] to change the gas
4.
concentration (ppm). Press 3[ENTER]to enter the
concentration level.
sensors together, or separately.
2
[
]
Yes
to initiate the calibration procedure.
[
]
BOTH
.
to exit calibration.
[OK]
or 3
Sensors
2
[Cancel]
to exit calibration.
. Press
2
[OK]. Enter the time

Advertisement