Download Print this page

Fluke 975CK Instruction Sheet page 4

Gas sensor calibration kit

Advertisement

2.
When zeroing is complete, remove neutral gas (Nitrogen) and
press 2[OK] or 3 [Cancel] to exit calibration mode.
Apply the CAL gas. Enter the amount using 1[DOWN]
3.
and 2[UP] to change the number and 3[ENTER] to
store the number.
Apply cal gas and press 2[OK] or 3 [Cancel] to exit
4.
calibration mode.
5.
The Meter finishes the procedure then prompts the user to
enter the time until the next calibration. Enter the amount
using 1[DOWN] and 2[UP] to change the number and
3[ENTER] to store the number.
CO2 sensor calibration is now complete and the Meter exits
calibration mode.
Contacting Fluke
To locate an authorized service center, visit us on the World
Wide Web at
following numbers:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
LIMITED WARRANTY AND LIMITATION OF LIABILITY
Regulators are warranted for 2 years from the date of purchase against
defects in the materials and assembly, and performance within specification.
If the regulator is considered by Fluke to be defective then Fluke will replace
the regulator at no charge. If the gas canister leaks or the gas becomes
unstable within 3 years from the manufacturing date on the canister, Fluke will
replace the canister at no charge. These warranties will not cover damage
from accident, neglect, misuse, alteration, contamination, or abnormal
conditions of operation or handling. Resellers are not authorized to extend any
other warranty on Fluke's behalf.
To obtain warranty service, contact your nearest Fluke authorized service
center to obtain return authorization information. Regulators need to be sent to
that Service Center with a description of the problem. Fluke will not accept
return of canisters. Customers must dispose of them in accordance with the
disposal instructions on the canister.
THIS WARRANTY IS BUYER'S ONLY REMEDY. NO OTHER WARRANTIES,
EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED
WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE, ARE GIVEN.
FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT,
INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING
FROM ANY CAUSE OR THEORY. Since some states or countries do not
allow the exclusion or limitation of an implied warranty or of incidental or
consequential damages, this limitation of liability may not apply to you.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
www.fluke.com
or call Fluke using any of the
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands

Advertisement

loading