Direct Current Verification Steps - Fluke 8808A Calibration Manual

Digital multimeter
Hide thumbs Also See for 8808A:
Table of Contents

Advertisement

8808A
INPUT
SENSE
V
4 W
2W/4W
HI
1000 V
CAT I
1000V
300V
MAX
600 V
CAT II
750V
1V
LO
500 V pk
200 mA
10 A
MAX
MAX
mA
NC
UUT
10A
Input
LO
Figure 3-5. 2 Amps and Greater Direct and Alternating Current Test Setup
Nominal
Range
Input
0 μA
200 μA
190 μA
200 μA
-190 μA
200 μA
0 μA
2000 μA
1900 μA
2000 μA
-1900 μA
2000 μA
0 mA
20 mA
19 mA
20 mA
-19 mA
20 mA
0 mA
200 mA
190 mA
200 mA
-190 mA
200 mA
Connect 57X0A and 5725A Callibrator to the UUT as follows:
HI
LO
10A
5725A
Current Output
Current Output
Table 3-6. Direct Current Verification Steps
90-Day Test Limit
High
0.010 μA
190.048 μA
-189.952 μA
0.10 μA
1900.39 μA
-1899.62 μA
0.0040 mA
19.0097 mA
-18.9903 mA
0.010 mA
190.048 mA
-189.952 mA
Performance Verification and Calibration
HI
HI
LO
LO
1-Year Test Limit
Low
High
-0.010 μA
0.010 μA
189.952 μA
190.067 μA
-190.048 μA
-189.933 μA
-0.10 μA
0.10 μA
1899.62 μA
1900.48 μA
-1900.39 μA
-1899.52 μA
-0.0040 mA
0.0040 mA
18.9903 mA
19.0116 mA
-19.0097 mA
-18.9884 mA
-0.010 mA
0.020 mA
189.952 mA
190.073 mA
190.048 mA
-189.927 mA
Direct Current Verification
57X0A
OUTPUT
SENSE
Ω
V A
VA
HI
WIDEBAND
HI
LO
LO
HI
AUX
GUARD
GROUND
CURRENT
5725A
CURRENT
OUTPUT
11A
MAX
2OV
PK
MAX
ffw005.eps
Low
-0.010 μA
189.933 μA
-190.067 μA
-0.10 μA
1899.52 μA
-1900.48 μA
-0.0040 mA
18.9884 mA
-190.0116 mA
-0.020 mA
189.927 mA
-190.073 mA
3
3-9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents