Commissioning
7.8 [2.09.5] Digital outputs
Table 7- 20
Function
(pre-selection)
Other functions
See also
Functions (Page 62)
68
Function (pre-selection) "Other functions"
Purpose/resulting settings
Set additional functions.
1. "Function" parameter field:
Authorize access with corresponding PIN, if necessary.
2. Select one of the offered functions:
–
Sync. (Function cannot currently be executed)
Synchronization of AutoCal sequences between several devices.
–
Calibration:
Total function check: signaling of "Calibration" operating state at device
level.
–
Calib. AM1/AM2
Signaling of "Calibration" operating state at module level.
–
Pump on/off
Switch external sample gas pump on/off.
–
Fct. check
Signal function check at device level.
–
Fct.check AM1/AM2
Signal function check at module level.
–
Signaling contact
Signaling contact: AutoCal signal contact, pulse 1000 ms.
Compact Operating Instructions, 01/2013, A5E31805656-01
Quick Start