4.2

Gas connections

4.2.1
Gas connections
NOTICE
Components in contact with sample gas
Components which come into contact with the sample gas must be resistant to it.
AM
OXYMAT 7 without internal reference
gas pump
Gas connections, analyzer module versions
Note
Reference gas connection
• The reference gas connection with PVDF screwed gland is only suitable for connecting
• The reference gas connection with ISO female thread is only suitable for connecting
Quick Start
Compact Operating Instructions, 01/2013, A5E31805656-01
an external reference gas pump, see module nameplate of analyzer module 7MB3020-
xAxxx-xxxx, letter A at 2nd position of order number.
reference gas pressures up to 5000 hPa, see module nameplate of analyzer module
7MB3020-xCxxx-xxxx, letter C, D or E at 2nd position of order number.
Enclosure
Rack-mounted device, rear
Wall-mounted device, bottom
Connecting
4.2 Gas connections
Gas connections
1 Sample gas inlet
2 Sample gas outlet
3 Not used
4 Reference gas inlet