To set the horizontal reference point

• Click on one of the arrows at the bottom of the waveform display.
The selected horizontal reference is highlighted.
The vertical arrows at the bottom of the display correspond to the left, center,
and right horizontal reference points. This is the position of the trigger if the
horizontal position is set to zero delay. A non-zero horizontal position value is
the time relative to the trigger event at the selected horizontal reference point.
Figure 3-34
Click to set the
horizontal reference
to the left side of the
display
Horizontal Reference Points
To set the horizontal reference point
Click to set the
horizontal reference
to the center of the
display
Using the Oscilloscope
Click to set the
horizontal reference
to the right side of the
display
3-45