Product Connections

The Product connects to infusion devices through the front-panel inlet ports. Fluid drain
hoses and accessories connections are made through the rear-panel connections.

Connect Infusion Devices

As shown in Figure 1, it is recommended that all infusion device connections be made to
the Product front-panel inlet connectors through 3-way stopcocks.
The channel 4 inlet shown in Figure 1, shows a 20 ml syringe attached to one 3-way
stopcock inlet. The syringe can be used to help priming. It can be used as shown or can
be connected further away from the inlet to help for flow tests. The syringe can be shared
among the channels and removed after the channel is primed.
Follow these recommendations when you connect to the inlet tubing circuits:
6
Figure 1. Infusion Device Connections to the Product
Use adequate prime volumes (for example, 10 ml) to push through any bubbles.
Use the stopcocks at the inlet to prevent fluid backflow out of inlets between tests.
When you connect to the inlet circuits (for example, when you attach the priming
syringes to the stopcocks) make sure no new bubbles are introduced.
20 ml Syringe
to aid priming
3-way Stopcock
From Infusion
Pumps
gir03.eps