Emerson Rosemount OCX8800 Instruction Manual page 208

Oxygen and combustibles transmitter, hazardous location probe with integral electronics
Hide thumbs Also See for Rosemount OCX8800:
Table of Contents

Advertisement

Instruction Manual
IM-106-880C, OI
4.
5.
6.
7.
8.
9.
10.
OCX with Integral
Electronics
Remove the cover from the electronics housing to expose the electronics housing
terminal blocks,
Figure
9-2.
Disconnect and remove the power leads from the AC power input terminal block and
remove the ground lead from the ground stud.
Disconnect and remove the O
terminal block.
If used, disconnect and remove the external relay leads from the alarm output relay
terminal block.
Disconnect and remove customer power and signal wire conduits and wiring from the
electronics housing.
Remove insulation to access the sensor housing mounting bolts. Unbolt the Hazardous
Area OCX 8800 from the stack and take it to a clean work area.
Allow the unit to cool to a comfortable working temperature.
and COe signal leads from the 4-20 mA signal output
2
OCX 8800 Removal and
Installation
April 2017
18
1