Emerson Rosemount OCX8800 Instruction Manual page 185

Oxygen and combustibles transmitter, hazardous location probe with integral electronics
Hide thumbs Also See for Rosemount OCX8800:
Table of Contents

Advertisement

Instruction Manual
IM-106-880C, OI
Index
Number
Units
DISABLE_2
16
DISABLE_3
17
DISABLE_4
18
GRANT_DENY
9
IN_1
11
IN_2
12
IN_3
13
IN_4
14
MIN_GOOD
20
OP_SELECT
22
OUT
7
OUT_UNITS
8
SELECTED
21
SELECT_TYPE
19
STATUS_OPTS
10
STRATEGY
3
1
TAG_DESC
2
23
Input Selector (ISEL) Function
Block
Description
None
A Connection from another block that disables the associated input fromthe
selection.
None
A Connection from another block that disables the associated input fromthe
selection.
None
A Connection from another block that disables the associated input fromthe
selection.
None
Options for controlling access of host computers and local control panels to
operating, tuning, and alarm parameters of the block. Not used by device.
Determinedby
The connection input from another block. One of the inputs to be selected
source
from.
Determinedby
The connection input from another block. One of the inputs to be selected
source
from.
Determinedby
The connection input from another block. One of the inputs to be selected
source
from.
Determinedby
The connection input from another block. One of the inputs to be selected
source
from.
None
The minimum number of good inputs The actual, target, permitted, and
normal modes of the block. Target: The mode to "go to" Actual: The mode the
"block is currently in" Permitted: Allow ed modes that target may take on
Normal: Most common mode for target.
None
Overrides the algorithm to select 1 of the 4 inputs regardless of the selection
type.
EU of IN
The block output value and status.
None
The engineering units of the output. Typically, all inputs have the same units
and the value is also the same.
None
The selected input number (1–4).
None
Specifies selection method (see Block Execution).
None
Allows selection of options for status handling and processing. The supported
status option for the PID block is Target to Manual if Bad IN.
None
The strategy field can be used to identify grouping of blocks. This data is not
checked or processed by the block.
None
The revision level of the static data associated w ith the function block. The
revision value w ill be incremented each time a static parameter value in the
block is changed.
None
The user description of the intended application of the block.
None
This alert is generated by any change to the static data.
Section 7: Foundation
Fieldbus
April 2017
15
9

Advertisement

Table of Contents
loading

Table of Contents