Adjust the concentration of chlorine until it is near the top end of the range. Press
ENTER.
The TIME DELAY message appears and will remain until the sensor reading is
relatively stable. To bypass the time delay, press ENTER. The GRAB SPL prompt
appears.
Take the sample of the process liquid and determine the concentration of chlorine in
the sample.
Press ENTER.
The PT2 prompt appears.
Use the arrow keys to change the flashing display to the concentration of chlorine
determined in the grab sample. Press ENTER to save.
Press RESET to return to the main display.