Switch-On Phase - T&D VEGAWELL 52 Operating Instructions Manual

Hydrostatic absolute pressure transmitter for gauge measurement in water and wastewater
Table of Contents

Advertisement

Connection via housing
Connection via VEGA-
DIS 62
VEGAWELL 52 • 4 ... 20 mA
Fig. 10: Wiring plan housing
To power supply or the processing system
1
1
Fig. 11: Wiring plan, VEGADIS 62
To the sensor
1
2
For power supply
Coupling for the connection cable to the indicating and adjustment module
3
Wire number
1
2

5.4 Switch-on phase

After connecting VEGAWELL 52 to power supply or after a voltage
recurrence, the instrument carries out a self-check:
Internal check of the electronics
l
4 ... 20 mA output jumps to the failure signal
l
5 Connecting to power supply
+
+
1
2
3
( )
( )
(-)
Wire colour/Polarity
brown (+)
blue (-)
Shielding
2
1
3
4
(-)
2
Terminal VEGADIS 62
1
2
Ground