Mounting Steps With Screwed Connection - T&D VEGAWELL 52 Operating Instructions Manual

Hydrostatic absolute pressure transmitter for gauge measurement in water and wastewater
Table of Contents

Advertisement

VEGAWELL 52 • 4 ... 20 mA

4.3 Mounting steps with screwed connection

Fig. 5: Threaded fitting
Suspension cable
1
2
Seal screw
Cone bushing
3
4
Seal cone
Threaded fitting
5
Seal
6
Mount VEGAWELL 52 with screwed connection as follows:
Weld the welded socket into the vessel top
1
Lower VEGAWELL 52 to the requested height by means on the
2
welded socket G1½ A or 1½ NPT on the vessel side
Insert the suspension cable from below into the open screwed
3
connection
Slide the seal cone and the cone sleeve over the suspension
4
cable, fasten manually with the seal screw
Screw the screwed connection into the socket, fasten with SW 30
5
and then fasten seal screw with SW 19
How to correct the height:
Loosen seal screw with SW 19
1
Slide seal cone and cone sleeve to the requested position on the
2
cable
Fasten the seal screw
3
Removal is carried out in reverse order.
4 Mounting
1
2
3
4
5
6
13

Advertisement

Table of Contents
loading

Related Products for T&D VEGAWELL 52

Table of Contents