Fluke 753 User Manual page 52

Documenting process calibrator
Hide thumbs Also See for 753:
Table of Contents

Advertisement

753/754
Users Manual
To measure temperature where an RTD input is used:
1.
If necessary, push  for MEASURE mode.
2.
Push t.
3.
Push  and  then Select RTD Type is shown.
4.
Push  or  to select the necessary RTD type.
5.
Push .
6.
Push  or  to select a 2-, 3-, or 4-wire connection.
The connections are shown on the display.
7.
Attach the RTD to input jacks as the display or Figure
14 shows. Use the supplied jumper between the mA Ω
RTD MEASURE low jack and the V MEASURE low jack
as shown if using a 3-wire connection.
8.
Push .
40
Shop for Fluke products online at:
To prevent possible damage to the Product, do
not force a dual banana plug between any two
jacks in the horizontal orientation. Doing so
will damage the jacks. Use the supplied jumper
wire when needed for RTD measurements. A
dual banana plug may be used in the vertical
orientation. See Figure 16.
MyFlukeStore
www.
Caution
1.888.610.7664
.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

754

Table of Contents