Connect a pressure module to the Product as shown in
Figure 14. The threads on the pressure modules accept
standard ¼ NPT pipe fittings. Use the supplied ¼ NPT to ¼
ISO adapter if necessary.
1.
Push for MEASURE mode.
2.
Push p. The Product automatically senses which
pressure module is attached and sets its range
accordingly.
3.
Zero the pressure. See the module's Instruction Sheet.
Modules can have different zeroing procedures that
depend on module type.
Note
Zeroing MUST be done before doing a task that
sources or measures pressure.
34
Shop for Fluke products online at:
4.
If necessary, the pressure display units can be changed
to psi, mHg, inHg, inH2O, ftH2O, mH2O, bar, Pa,
g/cm2, or inH2O@60°F. Metric units (kPa, mmHg, etc.)
are shown in Setup mode in their base units (Pa, mHg,
etc.). To change pressure display units:
1.
Push s.
2.
Push Next Page twice.
3.
Push or the Choices softkey with the cursor
on Pressure Units.
4.
Set the pressure units with or .
5.
Push .
6.
Push the Done softkey.