SITRANS F M
8.4.1

Basic settings

66
8. Commissioning
Decimal point for flow rate, totalizer 1 and totalizer 2 can be individually positioned.
open the respective window.
ensure that the cursor is positioned below the comma. Use the select key
move the comma to the requested position. Use the change key
Units are changed by means of the change key
Select units (cursor moved) by means of the select key
Totalizer 2 is not visible when batch is selected as digital output.
Q
2 - is only visible when it has been chosen as external input.
max.
Main frequency
Selects the main power supply frequency corresponding
to the country in which the flowmeter is installed.
(America = 60 Hz)
Flow direction
Selects the correct flow direction in the pipe
Q
max.
Sets the measuring range, the analog outputs and the
frequency output. Value, decimal point, unit and time can
be set individually (setting is dimension-dependent).
Q
2
max.
Sets the measuring range, the analog outputs and the
frequency output. Value, decimal point, unit and time can
be set individually (setting is dimension-dependent).
Only visible when it has been chosen as external digital
input.
Totalizers
Sets unit and decimal point.
Low flow cut off
Sets a % of selected Q
. To filter noise in the
max.
installation. Influences display and all outputs.
Error level
Selects which error level, the flowmeter will detect an error.
Empty pipe cut off
Set on - the alarm will indicate when sensor is
running empty. All readings, display and outputs will
indicate zero.
.
with the cursor placed below the unit selected.
.