Adding liquid detergents and neutralizers
^ Remove the container from the
washer.
^ Unscrew the siphon tube and remove
it.
^ Fill the container with the required
product.
^ Insert the siphon tube into the
container opening and screw it into
place.
Once the containers have been filled,
the message clears from the display.
30
Message "Check dispensing system
[X]"
The currently running program pauses.
^ Check the container(s) and
dispensing hoses indicated in the
display.
Refill or replace as needed.
Containers and dispensing systems
that are not in use can be shut off, to
avoid fault messages in the display
(see "Machine function / Query
container" in the Programming
manual).
Priming the dispensing system
Whenever a container has been
allowed to completely empty, it must be
primed (cleared of air) after refilling.
^ Select the corresponding service
program, e. g.,
^ Press the h start button.
DOS1-FILL
.