Calibration Adjustment - Fluke 233 Calibration Information Manual

True-rms remote display
Hide thumbs Also See for 233:
Table of Contents

Advertisement

233
Calibration Information
Step
Function
L
19
DC Volts
20
21
22
23
24
m
25
AC Millivolts
26
E
27
DC Millivolts
28
G Diode
29
E
31
Capacitance
32
35
36
Set calibrator to standby, reconfigure leads, and program for amps output
A
37
DC Amps
?
38
AC Amps
y
39
Temperature
40
41
[1] If using a Fluke 9100 calibrator, the Calibrator Frequency mode must be used to obtain accurate frequency.

Calibration Adjustment

The Meter features closed-case calibration adjustment using known reference sources. The Meter measures the
applied reference source, calculates correction factors, and stores the correction factors in nonvolatile memory.
The following sections present the features and Meter pushbutton functions available during the Calibration
Adjustment Procedure. Should the Meter fail any of the performance tests, perform the Calibration Adjustment
Procedure.
Use the following steps to view the Meter's calibration counter.
1. While pressing K, turn the rotary switch from OFF to Ω function. The Meter should display
"CAL".
2. Press g once to view the calibration counter. For example, "n001"
14
Table 3. Performance Tests (cont)
Range
Applied
6.000V
0V
6.000V
5V
60.00V
50V
600.0V
600V
600.0V
-600V
1000V
1000V
600.0 mV
6 mV, 45 Hz
600.0 mV
600mV, 1 kHz
600.0 mV
20 mV
600.0 mV
600mV
2.000 V
1.9V
1000 nF
30
9 μF
90 μF
9999 μF
900 μF
10.00 A
9.5 A
6.000 A
5.0 A, 45 Hz
Open input
0.0 °C
400 °C
Display Reading
-0.002 to 0.002
4.985 to 5.015
49.85 to 50.15
598.3 to 601.7
-598.3 to -601.7
995 to 1000
5.6 to 6.4
587.7 to 612.3
19.7 to 20.3
598.3 to 601.7
1.881 to 1.919
27 to 33
8.81 to 9.19
88.1 to 91.9
881 to 919
9.37 to 9.63
4.922 to 5.078
0PEn
-1.0 to 1.0
395.0 to 405.0