Calibration Adjustment; Preparing For Calibration Adjustment - Fluke ProcessMeter 789 Calibration Manual

Hide thumbs Also See for ProcessMeter 789:
Table of Contents

Advertisement

789
Calibration Manual

Calibration Adjustment

The Processmeter is calibrated using a closed-case procedure.
Calibrate the ProcessMeter once a year to ensure that it performs according to its
specifications.

Preparing for Calibration Adjustment

To avoid possible electric shock or personal injury:
24
Do not use the ProcessMeter if it looks damaged.
Inspect the ProcessMeter for damage, especially around the input
terminals. Inspect the test leads and test connections for damaged
insulation or exposed metal.
Look for cracks, missing plastic or damaged insulation. If damage
is detected, do not continue; contact Fluke to have the
ProcessMeter serviced.
Make sure that the battery compartment door on the ProcessMeter
is closed and latched before using it.
Check the test leads for continuity. Replace damaged test leads as
necessary.
Do not use the ProcessMeter if it appears to operate abnormally.
Protection designed into the ProcessMeter might be impaired. If in
doubt, have the ProcessMeter serviced.
To avoid electrical shock, always place the calibrator in the
Standby (STBY) mode between tests and before handling the test
connections or test cables.
Some of the calibration adjustment procedures involve the use of
high voltages and should be performed by qualified personnel
only.
WXWarning

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents