Installing The A70-2 Ppcl Pcb - Siemens SONOLINE G50 Service Manual

Ultrasound systems
Hide thumbs Also See for SONOLINE G50:
Table of Contents

Advertisement

Installing the G50 DIMAQ-IP
22. As demonstrated in Figure 5-5, attach the connector of the A70-2 PPCL
23. Install the four M3 X 6 (CK3 X 6) screw, supplied in the kit, through the A70-
24. Place the PPCL shielding plate on the host module, in its original position
25. As depicted in Figure 5-6, install the four screws that were removed from
26. Install connector J2 in the PPCL shielding plate and secure it in place with
!
Page 6 of 16
PCB to the A70 DVSC PCB located in the host module enclosure.
2 PPCL PCB, into the A70 DVSC PCB as indicated in Figure 5-5.
Figure 5-5 Installing the A70-2 PPCL PCB
(Figure 5-2).
the PPCL shielding plate shown in Figure 5-2.
the two M2.5 X 6 screws and external lock washers that are supplied in the
kit.
Caution: Use care to avoid dropping loose hardware into the system during the
disassembly and assembly of the G50. Loose hardware can short electronic
components. Safeguard all hardware for reinstallation.
7482255 Rev 01
October 2002
Siemens