Hach Orbisphere 3100 Basic User Manual page 20

Portable analyzer
Hide thumbs Also See for Orbisphere 3100:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
3. Sample temperature calibration
Option
Sample temperature
calibration
4. Miscellaneous
Option
Description
Enable raw
Check the box to enable the capture of raw data which is useful for troubleshooting
data logger
purposes
Activate
Check the box to activate the service timer option.
service timer
When activated, the instrument will automatically remind the user when the next sensor
service is due.
Enter the Nb of days between services in the box provided. This defines the due date of
the next sensor service.
If activated, select the Reset service timer option each time the instrument has been
serviced. This automatically sets the Last service date parameter to the current date.
Note: The total uptime of the instrument is displayed at the bottom of the screen.
Service O2 channel
1. O2: Calibration parameters
Option
O2: Calibration
parameters
2. O2: DC measurement parameters
Option
O2: DC measurement
parameters
3. O2: AC diagnostic parameters
Option
O2: AC diagnostic
parameters
4. O2: Calibration timer
Option
Description
Activate
Check the box to activate the calibration timer option.
calibration
When activated, the instrument will automatically remind the user when the next sensor
timer
calibration is due.
The Last calibration date parameter is shown for information. This is automatically
updated each time a calibration is performed on the sensor.
Enter the Nb of days between calibrations in the box provided. This defines the due
date of the next sensor calibration.
20 English
Description
You will need a sensor simulator for this option. Follow the on-screen
instructions to calibrate the sample temperature
Description
This option is useful for troubleshooting purposes and will display a number of
values associated with the measurement channel
Description
This option is useful for troubleshooting purposes and will display the
values of the fluorescent and reference LED current
Description
This option is useful for troubleshooting purposes and will display the values of
the fluorescent and reference amplitude and phase, plus the phase shift value