Performance Test Procedure - Fluke 333 Calibration Information Manual

33x series clamp meters
Hide thumbs Also See for 333:
Table of Contents

Advertisement

ClampMeters
Performance Tests

Performance Test Procedure

To test each of the Meter's functions and operating ranges, do the following:
1. Connect the source to the Meter's VΩ and COM input jacks.
2. Referring to Table 2, put the Meter in the desired function and range for each test.
3. Apply the indicated output from the source.
4. When using the amp function on the 5520A, make sure LCOMP on the 5520A is ON.
5. The reading on the Meter display should be within the low and high limits shown in the table.
6. Repeat steps 1-4 for each function and range in Table 2.
If the Meter fails to perform within the low-high range indicated for each test in Table 2, the Meter needs
to be calibrated and adjusted, or requires some repair.
50-Turn Current Coil
Conductor
5520A Calibrator
adc07f.eps
Figure 4. 33x Amps/Hz Verification Setup
7