9011 Dual-Well Calibrator
User's Guide
Controller Communications Commands continued
Command Description
Set ALPHA calibration parameter
to n
Read DELTA calibration parameter de[lta]
Set DELTA calibration parameter
to n
Read BETA calibration parameter
(cold side only)
Set BETA calibration parameter to
n (cold side only)
Miscellaneous (not on menus)
Read firmware version number
Read structure of all commands
Read all parameters
Legend:
Note:
Note:
42
Command
Command
Format
Example
al[pha]=n
al=0.0038433
de
de[lta]= n
de=1.45
be[ta]
be
be[ta]= n
be=0.342
*ver[sion]
*ver
h[elp]
h
all
all
[] Optional Command data
/ Alternate characters or data
{} Returns either information
n Numeric data supplied by user—may be entered in decimal or exponential notation
9 Numeric data returned to user
x Character data returned to user
When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by
a carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.
Commands that are preceded by a "C:" or an "H:" access the cold or hot side respectively. If the "C:" or
"H:" are not used, the command accesses the cold side.
Returned
de: 9.99999
be: 9.999
ver.9999,9.99
list of commands
list of parameters
and their values
Returned
Acceptable
Example
Values
0.0032 to
0.0042
de: 1.46126
1.0 to 1.9
be: 0.342
–25.0 to 25.0
ver.9011,1.11