Fluke 9011 User Manual page 43

High-accuracy dual-well calibrator
Table of Contents

Advertisement

9011 Dual-Well Calibrator
User's Guide
Table 2 Controller Communications Commands
Command Description
Display Temperature
Read current set-point
Set current set-point to n
Read temperature unit
Set temperature units:
Set temperature units to Celsius
Set temperature units to
Fahrenheit
Read scan function
Set scan function:
Turn scan function on
Turn scan function off
Read scan rate
Set scan rate to n degrees per
minute
Read display temperature hold
mode
Read hold status
Read temperature
Secondary Menu
Read proportional band setting
Set proportional band to n
Read cutout setting
Read heater power
(duty cycle)
Ramp and Soak Menu
Read number of programmable
set-points
Set number of programmable
set-points to n
Read programmable set-point
number n
Set programmable set-point num-
ber n to n
40
Command
Command
Format
Example
s[etpoint]
s
s[etpoint]=n
s=450
u[nits]
u
u[nits]=c/f
u[nits]=c
u=c
u[nits]=f
u=f
sc[an]
sc
sc[an]=on/of[f]
sc[an]=on
sc=on
sc[an]=of[f]
sc=of
sr[ate]
sr
sr[ate]= n
sr=5
hm[ode]
hm
ho[ld]
ho
t[emperature]
t
pr[op-band]
pr
pr[op-band]=n
pr=8.83
c[utout]
c
po[wer]
po
pn
pn
pn= n
pn=4
ps n
ps3
ps n = n
ps3=50
Returned
set: 9999.99 {C or
F}
u:{C or F}
scan: {ON or OFF} scan: ON
srat: 999.9 {C or
F}/min
hmode: AUTO
ho:{closed or
open}, 9999.9 {C or
F}
t: 9999.9 {C or F}
pb: 999.99999
cu: 9999 {C or F}
po: 9999.9
pn: 9
ps n : 9999.99 {C or
F}
Returned
Acceptable
Example
Values
set: 150.00 C
Controller
Range
u:C
C or F
ON or OFF
srat: 10.0
C/min
0.1 to 50°C
hmode: AUTO
ho: open,
75.0 C
t: 55.69 C
pb: 15.9
Depends on
Configuration
cu: 620 C, in
po: 1.0
pn: 2
2 to 8
ps1: 50.00 C
1 to 8, Instru-
ment Range