Figure 4-17. Signal During Simultaneous Sample-and-Hold Sampling

It is possible to enable and disable SS/H programmatically in NI-DAQmx,
although NI recommends that you leave SS/H enabled for most
applications. Disabling SS/H for one module disables this feature for all
modules in all chassis that are controlled by the same DAQ device. You
should only disable SS/H if your application does not require simultaneous
sampling and requires higher acquisition rates than are possible with SS/H
enabled. Refer to the
Developing Your Application in NI-DAQmx
of Chapter 5,
Using the
1
2
SCXI-1520, for more information about
4-29
Chapter 4
Theory of Operation
Time
3
section
SCXI-1520 User Manual