This document explains how to install, configure, and test the NI PXIe-5622 (NI 5622). The
NI 5622 is a 150 MS/s digitizer with onboard signal processing (OSP). The NI-SCOPE
instrument driver ships with the NI 5622 and programs the device.
Note
Before you begin, install and configure your chassis and controller.
To access NI 5622 documentation, navigate to Start»All Programs»National Instruments»
NI-SCOPE»Documentation.
Contents
Electromagnetic Compatibility Guidelines............................................................................... 1
Verifying the System Requirements..........................................................................................2
Unpacking the Kit..................................................................................................................... 2
Kit Contents...................................................................................................................... 3
Other Equipment....................................................................................................................... 3
Preparing the Environment....................................................................................................... 3
Installing the Software.............................................................................................................. 4
Installing the NI 5622............................................................................................................... 4
Hardware Front Panel Connectors.................................................................................... 6
Configuring the Hardware in MAX.......................................................................................... 7
Programming the NI 5622.........................................................................................................8
Troubleshooting........................................................................................................................ 9
Why Is the ACCESS LED Off When the Chassis is On?.................................................9
What Should I Do if the NI 5622 Doesn't Appear in MAX?............................................9
What Should I Do if the Module Fails the Self-Test?.....................................................10
Thermal Shutdown Error................................................................................................ 10
Where To Go Next.................................................................................................................. 11
Worldwide Support and Services............................................................................................ 11

Electromagnetic Compatibility Guidelines

This product was tested and complies with the regulatory requirements and limits for
electromagnetic compatibility (EMC) stated in the product specifications. These requirements
and limits provide reasonable protection against harmful interference when the product is
operated in the intended operational electromagnetic environment.

Advertisement

Table of Contents
loading

Summary of Contents for National Instruments PXIe-5622

  • Page 1: Table Of Contents

    NI PXIe-5622 16-bit 150 MS/s Digitizer This document explains how to install, configure, and test the NI PXIe-5622 (NI 5622). The NI 5622 is a 150 MS/s digitizer with onboard signal processing (OSP). The NI-SCOPE instrument driver ships with the NI 5622 and programs the device.
  • Page 2: Verifying The System Requirements

    Do not install a device if it appears damaged in any way. Unpack any other items and documentation from the kit. Store the device in the antistatic package when the device is not in use. 2 | ni.com | NI PXIe-5622 Getting Started Guide...
  • Page 3: Kit Contents

    Relative humidity range 10% to 90%, noncondensing (Tested in accordance with IEC-60068-2-56.) Maximum altitude 2,000 m (800 mbar) (at 25 °C ambient temperature) Pollution Degree Indoor use only. NI PXIe-5622 Getting Started Guide | © National Instruments | 3...
  • Page 4: Installing The Software

    Remove the black plastic connectors from all the captive screws on the module front panel. Identify a supported slot in the chassis. The following figure shows the symbols that indicate the slot types. 4 | ni.com | NI PXIe-5622 Getting Started Guide...
  • Page 5 11. Cover all empty slots using filler panels or slot blockers to maximize cooling air flow. 12. Power on the chassis. NI PXIe-5622 Getting Started Guide | © National Instruments | 5...
  • Page 6: Hardware Front Panel Connectors

    TTL (+5 V MAX) Table 1. NI 5622 Front Panel Signal Connections Connector Description Function IF IN SMA connector Analog input connection; digitizes data and triggers acquisitions PFI 1 SMB connector Digital trigger connection 6 | ni.com | NI PXIe-5622 Getting Started Guide...
  • Page 7: Configuring The Hardware In Max

    Compensation from the SFP menu. Note The NI 5622 has self-calibration capabilities, which you can access programmatically with NI-SCOPE and your ADE, or interactively with NI-SCOPE SFP or MAX. NI PXIe-5622 Getting Started Guide | © National Instruments | 7...
  • Page 8: Programming The Ni 5622

    \Public\Documents topic of the NI High-Speed \National Digitizers Help to manually add all Instruments required include and library files to • Windows XP— Documents your project. and Settings\All Users\Shared Documents\National Instruments 8 | ni.com | NI PXIe-5622 Getting Started Guide...
  • Page 9: Troubleshooting

    Windows 8 Right-click the Start screen, and select All apps»Control Panel» Hardware and Sound»Device Manager. Windows 7 Select Start»Control Panel»Device Manager. Windows Vista Select Start»Control Panel»System and Maintenance»Device Manager. NI PXIe-5622 Getting Started Guide | © National Instruments | 9...
  • Page 10: What Should I Do If The Module Fails The Self-Test

    Select Start»Control Panel»System»Hardware»Device Manager. If you are using a PXI or PXI Express controller, verify that a National Instruments entry appears in the system device list. Reinstall NI-SCOPE and the device if error conditions appear in the list. If you are using an MXI controller, right-click PCI-to-PCI Bridge, and select Properties from the shortcut menu to verify that the bridge is enabled.
  • Page 11: Where To Go Next

    Located online at ni.com/manuals Located using the NI Example Finder Worldwide Support and Services The National Instruments website is your complete resource for technical support. At ni.com/ support, you have access to everything from troubleshooting and application development self-help resources to email and phone assistance from NI Application Engineers.
  • Page 12 Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the file on your media, or the National Instruments Patent Notice at .