Fluke 753 User Manual page 90

Documenting process calibrator
Hide thumbs Also See for 753:
Table of Contents

Advertisement

753/754
Users Manual
16. Push  to move through the choices to Set Trip
Function to Trip Cont.
17. Push the Done softkey.
18. Push the Manual Test softkey.
19. Close the pressure line vent and slowly move the
pressure up to the trip point.
20. When the switch sets, slowly decrease the pressure
until the switch resets. If necessary, this cycle can be
done again.
21. Push the Done softkey and see the results.
22. Push the Done softkey and if necessary, record Tag,
S/N, and/or ID.
23. Push the Done softkey.
24. Exercise the switch by varying the applied pressure.
Adjust the switch until the set point is correct.
25. Use the softkeys to control the Product, and adjust the
switch as necessary.
78
info@Fluke-Direct.ca
26. Push the Done softkey.
27. Push the As Left softkey to start the test again with the
same parameters. Results from the As Found and As
Left tests are kept in Product memory to view later or
upload.
The procedure for switches that respond to other
parameters work similarly. When you do a 2 Pt. Switch Test,
follow the directions given on the display for the first switch
test, change test leads, and do the second switch test.