17. Push the Done softkey to accept the calibration
parameters. The display changes to the screen shown
in Figure 32.
Figure 32. Measure and Source Screen for Calibration
18. At this time you can do an automatic test or step
through the test points manually. Push the Auto Test
softkey to have the Product go through the tests
automatically. If necessary, push Abort to exit from the
calibration procedure. The tests start at the first test
point, sources the correct temperature and measures
the corresponding current from the transmitter.
info@Fluke-Direct.ca
When a measurement is stable and recorded, the
Product goes to the subsequent step. Because the
Product waits until the measurement becomes stable,
the Auto Test works as necessary for instruments with
built-in damping. The error of the expected measured
value is shown in the top left of the measure window.
19. The Product moves to the remaining set of points. For
temperature and electrical parameter calibration, the
points are done automatically. If your source pressure,
the Product stops at each step to let you adjust the
pressure source. When the tests are complete, an error
summary table similar to Figure 33 is shown.
gks46s.bmp
Fluke-Direct
.ca
Documenting Process Calibrator
Process Instrument Calibration
Figure 33. Error Summary Screen