To measure temperature where an RTD input is used:
1.
If necessary, push  for MEASURE mode.
2.
Push t.
3.
Push  and  then Select RTD Type is shown.
4.
Push  or  to select the necessary RTD type.
5.
Push .
6.
Push  or  to select a 2-, 3-, or 4-wire connection.
The connections are shown on the display.
7.
Attach the RTD to input jacks as the display or Figure
14 shows. Use the supplied jumper between the mA Ω
RTD MEASURE low jack and the V MEASURE low jack
as shown if using a 3-wire connection.
8.
Push .