Chapter 9 Specifications - Omega Engineering OMEGAFLEX FPU5-MT-110 User Manual

Peristaltic pump motor
Table of Contents

Advertisement

Specifications
9
Maximum Fluid Back Pressure:
Max Suction Lift:
Temperature Range (Fluid):
Flow Rate:
Motor:
Display:
Power:
Line cord:
Housing:
Dimensions (H x W x D):
Weight:
Fuse:
Tubing Wall
Tubing Size
Thickness
Inner Diameter
"
"
1
1
1 6
3 2
"
"
1
1
1 6
1 6
"
"
1
1
1 6
8
"
"
1
3
1 6
1 6
"
"
1
1
1 6
4
"
"
1
5
1 6
1 6
Table 9-1. Calibration Accuracy
Calibration
Accuracy
±4%
±2%
±2%
±2%
±2%
±4%
Calibration accuracy studies were done by calibrating the pump motor and
comparing actual volume pumped to requested total volume. Various tube
materials were used along with 20°C water, 0 PSI back pressure.
Accuracies may vary with fluid viscosity, temperature and back pressure.
20 PSI
20 feet of H
O (6.1 meters of H
2
-50°F to 300°F (-46°C to -149°C)
1 to 2280 mL/min
Reversible, 75 W (
1
1 0
600 rpm)
16 character dot matrix LCD, 9 mm (
characters
90 to 130 Vac, 1.5 A; 190 to 260 Vac, 0.8 A
(Unit may be used with extension cord)
1.8 m (6 ft) with US standard plug on 115 Vac
units 1.8 m (6 ft) with European plug on
230 Vac units
Painted sheet metal with rubber feet; units are
stackable
152 x 254 x 254 mm (6" x 10" x 10")
8.2 kg (18 lb)
2 amp, 110 Vac or 1 amp, 220 Vac
Tubing Wall
Tubing Size
Thickness
Inner Diameter
1.5 mm
1.0 mm
1.5 mm
2.0 mm
1.5 mm
3.0 mm
1.5 mm
4.0 mm
1.5 mm
5.0 mm
1.5 mm
6.0 mm
1.5 mm
7.0 mm
1.5 mm
8.0 mm
NOTE
O)
2
HP), 90 Vdc (10 to
") high
3
8
Calibration
Accuracy
±4%
±2%
±2%
±2%
±2%
±2%
±4%
±4%
9-1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fpu5-mt-220

Table of Contents