Operating the Pump Motor in the STOPPED Display Mode

5.4 Priming the Tubing

Pressing the PRIME key causes the tubing to fill with fluid.
1.
2.
Extensive testing has shown that the minimum motor speed required to prime
the tubing varies significantly with the size of the tubing. These variances are
shown in Table 9-3 in Chapter 9. If the tubing will not prime regardless of motor
speed, simply press on the pump stator (refer to Figure 3-11) while the stator is
shut and the motor is running. Pressing on the stator enhances the priming
action of the pump. Release pressure after the tubing is primed.

5.5 Calibrating the Tubing

CALIBRATE
Display shows
Prime or START?
ACTION
Press PRIME to
prime tube or
START to dispense
desired amount of
fluid
5-6
FLOW RATE setting
To prime the tubing, press and hold down the PRIME key. As long as you hold
down the PRIME key, fluid flows in the tubing, and the display shows
"Priming...".
To stop priming a tube, simply release the PRIME key. The pump motor stops
running, and fluid stops flowing.
PRIME
Display shows
Priming...
ACTION
Hold button down to dispense fluid
START
STOP
Display shows
Dispensing . . .
ACTION
Press Start/Stop button to
dispense fluid. Press again
to stop the process.
PRIME
Display shows
Priming. . .
ACTION
Pumps fluid at
current RPM or
as long as button is
pressed
Display shows
xxx mL or xxx Liters
ACTION
Scrolls up. Enter TOTAL VOLUME dispensed
CALIBRATE
Display shows
CALIBRATED!
ACTION
Red LED next to
Calibrate key
lights up