14. Calibration Procedure for HX93AV (all styles)

Relative Humidity Adjustment

Recommended equipment:
• Humidity Calibration Kit, OMEGA
• Handheld Digital Multi-meter, OMEGA
• Dc Power Supply, OMEGA
1. Remove enclosure cover.
2. Connect transmitter as shown in figure below.
3. Apply power to transmitter and allow to warm up for 15 min.
4. Place unit in a 11% RH environment and allow to stabilize for 15 min.
5. Adjust potentiometer "P3" for a output reading of .110 Vdc.
6. Place unit in a 75% RH environment and allow to stabilize for 15 min.
7. Adjust potentiometer "P2" for a output reading of .750 Vdc.
8. Repeat steps 4, 5, 6, and 7 as necessary until proper readings are
maintained.
9. Calibration complete.