SECTION 22—TIMING
10. Turn the unit off and then back on. The timing value is stored in memory for future 
use. 
11. If the unit is turned off before an autotune cycle is completed, the P.E. valve timing will 
not be stored or updated. If the board is new, autotuning MUST be successfully 
completed before the unit can be placed into service.
12. After a successful autotune, run the unit for 30 minutes and verify that the oxygen 
concentration output of the unit is within specification.
13. If the oxygen output concentration is not within specification, rerun the autotune 
again.
14. Reinstall the cover screws.
NOTE: PC board is exploded 
away from control panel for 
clarity. there is no need to 
remove P.C. Board from 
control panel.
Perfecto
V
2
P.C. Board
FIGURE 22.1 Autotuning
74
Autotune Push Button Switch