Backup Cooling Kit (Option) - Panasonic MDF-U4186S Series Operating Instructions Manual

Ultra-low temperature freezer
Table of Contents

Advertisement

BACKUP COOLING KIT (OPTION)

WARNING
As with any equipment that uses CO
of the equipment. It is important that you assess the work site to endure there is suitable and
sufficient ventilation. If restricted ventilation is suspected, then other methods of ensuring a safe
environment must be considered.
devices.
This freezer can be provided with a backup cooling kit (CVK-UB2) which is available as an optional
component. For the installation, refer to the instruction manual enclosed with the system.
1. Switch of backup cooling kit (BACKUP)
When turning on the system, the lamp is brightened.
This means that the system is ready.
operation of the system, turn off this switch.
2. Test switch (TEST)
This switch is for checking the operation of backup
cooling kit.
Pressing this switch is resulted in the
release of liquid carbon dioxide without system
operation.
3. Temperature setting knob (TEMP. SET)
With this knob, set the temperature at which the system
is operated.
The effective set temperature range is
o
between -50
C and -70
gas, there is a likelihood of oxygen depletion in the vicinity
2
These may include atmosphere monitoring and warning
To stop the
o
C.
31
1
2
3