4. Taking medication

1.
Hold the nebuliser kit as indicated on the right.
Caution:
Do not tilt the nebuliser kit at an
angle of greater than 45° in all
direction. Medication may flow
into the mouth.
2.
Press the power switch to the on (
The compressor starts and nebulisation starts.
Note: Observe whether aerosol is generated.
To stop nebulisation
Press the power switch to the off (
Warning:
Do not cover the compressor with a blanket or towel etc., during use.
Caution:
• Do not block the air filter cover.
• Limit use to 20 minutes at a time, and allow a 40 minute interval
before using the device again.
• When using the device, the main unit may become hot.
• Do not touch the main unit for other than necessary operation such
as turning off the power while nebulising.
Note: Do not leave the compressor on for a long period of time. This could
result in the compressor overheating or malfunctioning.