Physical
Dimensions (H x W x D):
Air outlet (complies with ISO 5356-1:2004):
Weight:
Housing construction:
Air filter
Standard:
Hypoallergenic:
Aircraft use
ResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21,
category M) for all phases of air travel.
Wireless module
Technology used:
It is recommended that the device is a minimum distance of 2 cm from the body during operation. Not applicable to masks,
tubes or accessories
Operating pressure range
S, VAuto:
CPAP:
Supplemental oxygen
Maximum flow:
Pneumatic flow path
Design life
Device, power supply unit:
Cleanable humidifier:
Air tubing:
General
The patient is an intended operator.
Air tubing
Air tubing
Material
ClimateLineAir
Flexible plastic and electrical components
SlimLine
Flexible plastic
Standard
Flexible plastic
Heated air tubing temperature cut-out:  41°C
116 mm x 205 mm x 150 mm
22 mm
1115 g
Flame retardant engineering thermoplastic
Material: Polyester non woven fibre
Average arrestance: >75% for ~7 micron dust
Material: Acrylic and polypropylene fibres in a polypropylene
carrier
Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron
dust
2G GSM
2 to 25 cm H
O (2 to 25 hPa)
2
4 to 20 cm H
O (4 to 20 hPa)
2
15 L/min (CPAP, S); 4 L/min (VAuto)
1. Flow sensor
2. Blower
3. Pressure sensor
4. Mask
5. Air tubing
6. Side cover
7. Device
8. Inlet filter
5 years
2.5 years
6 months
Length
2 m
1.8 m
2 m
Inner diameter
15 mm
15 mm
19 mm
English