Program settings
Program header
84
Adjusting program settings
The program settings should be adjusted to suit technical
requirements and the load.
Additional in-depth knowledge is required to alter program settings
and this should only be done by an experienced user or by a Miele
Service Technician.
Program and dispensing changes must be documented for
validated processes according to local regulations.
The processes must be validated again as necessary.
Program structure
Each program is subdivided into program blocks which run one after
another. A program consists of at least one and a maximum of 11
program blocks. Each block can occur only once in a program.
The so-called program header is placed above the program blocks
and contains general program settings. Individual wash block
parameters are also globally activated or deactivated here.
Water volume change
–
Each program block is allocated a nominal water quantity. The
water quantity can be raised or lowered incrementally for all blocks
to the base value in the program overview.
–
Drainage time
If the on-site waste water system is insufficient to drain the waste
water from the wash cabinet within the time allocated, the drainage
time can be lengthened by a set amount.
Parameters for measuring water pressure and spray arm pressure
can only be accessed by Miele Professional Service.