Application technology
Test point for
measuring water
pressure
36

Wash pressure measurement

The water pressure can be measured on all mobile units and baskets
with spray arms, injector bars or other wash connections, e.g. during
performance tests and validations in accordance with EN ISO 15883.
On mobile units and baskets with spray arms and additional injector
bars or other wash connections, there is a connection on the injector
bar or a wash connection for water pressure measurement. The exact
location is described in the respective operating instructions for the
mobile unit / baskets.
On mobile units and baskets with spray arms and no additional wash
connections, access for the wash pressure measurement is provided
on the side of the water supply pipe.
To measure water pressure, replace the blind stopper with a Luer
Lock adapter, e.g. E 447.
Under no circumstances may items to be washed, washing
attachments etc. be connected to the test point. After the
measurement, the test point must be closed again with the blind
stopper.