Switching monitors

Wear one of the monitors while the other one remains in the charging cradle. Switch the monitor on your
chest in the morning, at night before you go to sleep, or during the day if the battery is low.
Remove
You can switch the monitor without removing
the Strip from your body. No action needs to
be performed on the smartphone.
1. Hold the Strip against your skin.
2. Lift one side of the monitor until the snaps
release. Repeat on the other side.
3. Snap the monitor into the plugged-in
charging cradle.

Removing the Strip

Change the Strip every three days OR when it no longer adheres to the skin. Strips are disposable.
1. Remove the Strip while showering or use a warm, damp cloth to blot while removing.
2. Slowly peel back and roll the adhesive away from your chest, continually blotting with water.
Important: Do not rip or tear of the Strip quickly. Improper removal may cause mild skin irritation.
Attach
Remain near the smartphone when attaching
the fully-charged monitor to the Strip.
1. Snap fully-charged monitor to Strip.
2. Power on the monitor by pressing and
releasing the center button. Do not hold the
button down.
3. The smartphone will automatically detect
the new monitor.