pH Operations
12. After reviewing the calibration data, push PH or STORE to begin reading the
pH value of any solution. If the meter is setup for three or more calibration
points, continue with the next step.
If setup is for a 3 point calibration:.
1.
Push ENTER. If OK, push the arrow keys to change the value then press the
pH to accept just a 2 point calibration and display the pH measurement mode.
2.
If ENTER is selected in the set up for additional calibration point, CAL 3 will
stop flashing and the buffer value will start flashing until calibration is complete.
3.
When done with the final calibration point, calibration is complete. The meter
will beep three times and the meter is ready to begin reading the pH value of
any solution.
28