Section 1 Specifications Specifications are subject to change without notice. Meter specification Details Dimensions 3.5” W x 8” H x 2” D (90mm x 204mm x 50mm) Weight 3.0 lbs (1.3 Kg) Battery requirements 4-ANSI -ISA or IEC-LR6 (AA Alkaline) Power consumptions 1W backlight on Battery life...
Page 6
Specifications Conductivity Calibration Up to five points Accuracy ± 1% full scale or ±1st digit Resolution 0.01 ìS, 0.1 ìS ,1 ìS, 0.01 mS, 0.1 mS Autoranging 0.00 to 19.99 ìS, 20.0 to 199.9 ìS, 200 to 1999 ìS, Range 2.00 to 19.99 mS, 20.0 to 199.9 mS Calibration Up to five points...
Section 2 General information In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential damages resulting from any defect or omission in this manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notices or obligation.
The meter automatically identifies which type of electrode is attached when the meter is turned on The H160 and H170 meters accept both traditional style glass pH with BNC connectors and non-glass pH probes with ISFET (Ion Sensitive Field Effect Transistor) silicon chip sensors.
Installation 3.2 Probe connection installation N O T I C E The ISFET probe connector and conductivity probe connector are keyed differently to prevent attachment of a probe to the wrong connector. The meters are designed to use only one type of pH probe at a time. If the meter is submerged, and the rubber plug is not installed on the 3.5 mm phono jack, the interior of the jack may fill with water although the interior of the meter itself remains sealed.
Always turn off the meter when a probe is changed. Connect the BNC probe temperature sensor to the 3.5 mm phono jack or BNC connection Both the H160 and H170 use a 3.5 mm phono jack connector and a BNC connector. Refer to Section 3 Installation on page 7 to view the connections.
Section 4 Operation 4.1 Keypad description Refer to Figure 3 for keypad and key descriptions and functions. Figure 3 Model H170 shown Action H160 H170 • • Turn meter on and off • • Display pH • • Display mV •...
Page 14
Operation Action H160 H170 HOLD mode. Readings are frozen until HOLD key is • • pressed again to release the display. Only operates in measurement modes. • • • Used to confirm values and begin calibration Scroll values when in calibration mode: View setup screens •...
Operation 4.2 Display description The H160 and H170 Waterproof Handheld meters each display features in an LCD display as follows: Figure 4 LCD display and features Item Description H160 H170 • • ISFET pH probe indicator • • The meter is in the pH measurement mode •...
Operation Item Description H160 H170 • TDS or salinity mode value is in mg/L or g/L • TDS or salinity value is ppm • TDS or salinity value is in ppt • • The displayed value is a mV value •...
Operation To set the stabilization lock: Push the SETUP key to put in the setup mode. Push ENTER. The display will show the stabilization lock status. Use the arrows keys to set the stabilization lock mode to off or on. Push ENTER to save changes and return to setup mode.
Operation Push the arrow keys until the temperature units setup screen is displayed. Push ENTER. The display will show the current units. Use the arrow keys to set the temperature units. Push ENTER to save changes and return to setup mode or push SETUP to cancel any changes and return to the setup mode.
Operation Push the arrow keys until the sound icon is displayed. Push ENTER. The display will show the sound status (ON or OFF). Use the arrow keys to display ON or OFF. Push ENTER to save changes and return to setup mode. Push SETUP to return to setup mode without saving changes.
Operation To change or set the meter shutoff time: Push the SETUP key to put in the setup mode. Push the arrow keys until the meter automatic shutoff time screen is displayed. Push ENTER. The display will show the number of minutes until the meter automatically shuts off.
Page 21
Operation Push the arrow keys until the clear all calibration points screen is displayed. Push ENTER. The meter will prompt if the user wants to clear all calibration points. The default is No. Push the arrow keys to display Yes and then Push ENTER to clear. Push SETUP to return to setup mode without saving changes.
Operation 4.4 ERROR CODES N O T I C E It is not recommended that the user override the error codes. It is not possible to override all error codes. Pushing the arrow keys will display the Yes and NO options. Some error codes, such as poor electrode condition, can be overridden by the user to allow the instrument to be used until a replacement electrode can be obtained.
Section 5 pH Operations 5.1 Set the pH options 5.1.1 Set the pH resolution: 0.1 pH or 0.01 pH The pH can be displayed in two resolutions: 0.1 pH or 0.01 pH. The stabilization lock (READY icon) will occur faster at 0.1 pH resolution. To select the resolution: Push SETUP then push PH to put the meter in the pH setup mode.
pH Operations Push the arrow keys until the pH calibration point setup screen is displayed. Push ENTER. The display will show the current number of calibration points. Use the arrow keys to set the calibration points. Push ENTER to save changes and return to setup mode. Push SETUP to return to setup mode without saving changes.
Page 25
pH Operations Push ENTER. The display will show the pH buffer for the first calibration point. Push ENTER to set the first calibration buffer. The display will show the pH buffer for the next calibration point. After pushing ENTER on the last calibration buffer, the change will be saved and the meter will return to the setup mode.
pH Operations 5.2 Set the manual temperature compensation for glass pH electrodes Glass pH electrodes are automatically temperature compensated if a temperature probe is detected in the 3.5 mm phono jack. If no probe is detected, the system defaults to manual temperature compensation at 25.0 C. Manual temperature compensation is always in °C (No °F conversion is necessary).
Push ENTER to save changes and return to setup mode. Push SETUP to return to setup mode without saving changes. 5.3 pH calibration The H160 and H170 meters accept one, two, or three calibration points. The meter defaults to a 2-point calibration. To calibrate the pH sensor: Push PH to put in the pH mode.
Page 28
pH Operations Push CAL. The CAL icon will flash. If the buffer is correct, push ENTER. The pH buffer value will flash until calibration is complete. If the pH buffer is incorrect, push the arrow keys to select another buffer. Repeatedly pushing or will display the choices of availability buffers 1.68, 4.01.
Page 29
pH Operations When calibration is complete for that buffer, the meter will beep three times, and the display will show a flashing CAL 2 and the next pH buffer value factory default is 4.01). Rinse the probe in deionized water and place in second buffer solution. Push ENTER if OK, or to change value.
Page 30
pH Operations 12. After reviewing the calibration data, push PH or STORE to begin reading the pH value of any solution. If the meter is setup for three or more calibration points, continue with the next step. If setup is for a 3 point calibration:. Push ENTER.
Section 6 Conductivity & TDS Operations 6.1 Set the Conductivity options 6.1.1 Set the conductivity calibration points The meter stores up to five calibration points. Stored conductivity calibration points can be viewed, and are displayed in order of lowest to highest conductivity value. Push SETUP then push COND to put in the conductivity setup mode.
Conductivity & TDS Operations Push the arrow keys until the conductivity temperature normalization screen is displayed. Push ENTER. The display will show the conductivity normalization value. Use the display settings between 20.0 °C and 25.0 °C. Push ENTER to save changes and return to setup mode. Push SETUP to return to setup mode without saving changes.
Conductivity & TDS Operations Push the arrow keys until the conductivity probe constant screen is displayed. Push ENTER. 6.2 Set the TDS options 6.2.1 Set the TDS conversion factor TDS values are related to conductivity. The operator can calibrate using a TDS standard or calibrate using conductivity standards then program the meter with a conversion factor.
Conductivity & TDS Operations Push ENTER. The display will show the current conversion factor. Use the arrow keys to set the conversion factor. Push ENTER to save changes and return to setup mode. Push SETUP to return to setup mode without saving changes. 6.2.2 Set TDS units TDS can be displayed in metric units mg/L (g/L) or ppm (ppt).
Conductivity & TDS Operations 6.2.3 Set salinity units: ppm (ppt) or % Push SETUP then SAL to put in the salinity setup mode. Push the arrow keys until the salinity units screen is displayed. Push ENTER. The display will show the salinity units. Use the arrow keys to set the salinity units (ppt or %).
Page 36
Conductivity & TDS Operations temperature compensation coefficient of 2% per °C. The coefficient is user selectable in the setup mode. Note: The meter automatically detects the cell constant. Calibration is automatically cleared if the conductivity probe is changed to a probe of a different cell constant. Push COND or push TDS.
Page 37
Conductivity & TDS Operations Upon completion of the calibration, the meter will display the slope as compared to the nominal sensitivity for a probe of that cell constant and the total number of conductivity and TDS calibration points stored in the meter. Note: Number of calibration points in the meter.
Section 7 Maintenance 7.1 ISFET pH probes maintenance D A N G E R Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury. Note: The following information applies to new ISFSET pH Probes and/or sensors left in extended dry storage.
Maintenance • Clean any reference gel off the probe sensor AND out of the rubber dust cap’s inside surface. • Soak the new probes or probes that have been stored for an extended time for at least 10 minutes in pH 7 prior to use. •...
Maintenance 7.4 Probe cleaning N O T I C E Do not use metal objects such as pins or paperclips to clean the sensor. The sensor can be permanently damaged with aggressive abrasion of the sensor surface. Figure 5 Probe cleaning 7.4.1 ISFET non-glass probe cleaning and reconditioning N O T I C E...
Maintenance 7.4.1.2 Recondition the ISFET non-glass probes N O T I C E Do not use the probe as a thermometer to determine if the buffer is above 60 °C. Sudden immersion in excessively hot fluids may permanently damage the probe. ISFET probes are designed to be stored dry and have a virtually unlimited shelf life;...
Section 8 Troubleshooting 8.1 Error codes Indication Cause User Actions Auto power off has No display Push the power button. cleared display. Follow the probe cleaning procedures described in Dirty probe. this manual. Clean the probe contacts on the probe cable connector and on the meter with methanol and a Dirty probe/meter cotton swab and let dry completely.
Page 44
Troubleshooting Indication Cause User Actions Presence of the ISFET probe is not Turn on the meter. If the ISFET icon is not displayed, The meter detected by the replace the ISFET probe. continually meter or the probe. displays Carefully follow the two point calibration instructions. 14.00 or 0.00 Out of calibration.
Troubleshooting 8.1.1 Low battery indicator A low battery icon will appear when approximately 25 hours of battery life remains. If the batteries are not replaced after the low battery icon appears, the batteries will eventually be depleted to the level where the instrument cannot reliably be used without the possibility of measurement errors.
HACH Company World Repair Service in the Repair Service in Headquarters United States: Canada: P.O. Box 389 HACH Company Hach Sales & Service Loveland, Colorado Ames Service Canada Ltd. 80539-0389 U.S.A. 100 Dayton Avenue 1313 Border Street, Unit 34 Tel (800) 227-HACH...
Page 50
Contact Information HACH LANGE S.L.U. HACH LANGE LDA HACH LANGE SP.ZO.O. Edif. Arteaga Centrum Av. do Forte nº8 ul. Opolska 143 a C/Larrauri, 1C- 2ª Pl. Fracção M PL-52-013 Wrocław E-48160 Derio/Vizcaya P-2790-072 Carnaxide Tel. +48 (0)71 342 10-83 Tel. +34 94 657 33 88 Tel.
Page 51
Index ......17 automatic shuttoff ..24 set manual temperature ......21 set resolution power batteries ......14 low indicator ....43 low battery indicator salinity ........37 Cleaning ........ 33 set units cleaning sound ........38 probes ........16 on/off compensation controls and settings stablilzation lock conductivity .........