Cover Replacement; Fan Filter Cleaning - Keithley 3930A Service Manual

Multifunction synthesizer
Hide thumbs Also See for 3930A:
Table of Contents

Advertisement

SECTION3
Service Information
Model
393019
~"l.4tvwcnm~"
Model 7755
5OC2
Feed Through
Function
output
Model 7051-2
I
Coaxial Cable
?gure 3-7.
Connections to DVM
Display Contrast
1. Press the front panel DSM, key to return the display
to normal.
2. Adjust R752 (CONTRAST) for the desired display
contrast.
3.3.8
Cover Replacement
After calibration, replace the top and bottom covers, and
secure them with the four screws removed earlier. Be
careful not to scratch the conductive coating applied to
the inside of the covers, and be sure not to peel off the
front panel polyester film.
3.4
FAN FILTER CLEANING
The fan filter should be cleaned at least once every three
months when the unit is operated in a clean environment
or at least once a month when the unit is operated in a
dirty environment. The fan filter element should be
cleaned as follows:
1. Turn off the instrument power and disconnect the
3. Set J401 (FCTN OUT) to the desired position (see
line cord.
Figure 3-3 for location). Set the jumper to FCTN OUT
2. Remove the filter cover on the rear panel.
ON to enable FCTN OUT at power on, or set it to
3. Remove the filter element from the filter cover.
FCTN OUT OFF to leave FCTN OUT disabled at
4. Soak the filter element in a solution of mild detergent
power on.
and water until clean.
4. Replace the top and bottom covers.
5. Rinse the filter element thoroughly in clean water,
then allow the filter to dry thoroughly before re-
placement.
6. When the filter has dried completely, replace the
cover and filter assembly.
CAUTION
The instrument
should not be operated
without the filter in place.
3.5
FCTN OUT JUMPER
The FCTN OUT jumper, which is located on the control
board, allows you to select whether or not the output sig-
nal is turned on when power is first applied to the instru-
ment. (As shipped, FCTN OUT is enabled when power is
turned on). Use the procedure below to set the jumper
position.
1. Disconnect the line cord and all other instruments
from the Model 3930A.
2. Remove the top and bottom covers (see paragraph
3.3.5 for details).
3-9