8. CALIBRATION
Figure 8-10. XCALMON ITP Starting Window (Using Production Loader)

8.4.3 Calibration Procedure

Calibration procedure of XCALMON was the same as below procedure.
- Configuration of calibration
- Running ITP using production loader
- Calibration start using XCALMON
- Verification of calibration result
A. Configuration of Calibration
Configure to calibrated U8130 mobile phone like Figure7-1. If configuration will be accomplished, start
XCALMON program.
B. Running ITP Using Production Loader
If XCALMON will be executed, you should run ITP using "L" ITP starting icon at first.
Click the "L" icon, then you will see the ITP start window like Figure7-10.
When you will turn on the U8130 mobile phone, the production loader will be downloaded
automatically like Figure7-11 and then it will execute the ITP at once.
If the ITP will operate normally, you should see the characters "TP, OK" in ITP command window like
Figure7-12.